[. . . ] -EN- CSK8B - CSK8B – USER MANUAL Read the safety warnings and instructions for the thermodisinfectors in which the rack is used with care, together with the safety instructions of the producer of the glassware and any other item placed on the rack. [. . . ] You can see that the instrument blocks the spray arms, with consequent bad washing In the near picture, you can see a correct positioning of surgical instrument, that don’t stop the sprayarms 19390404000_CSK8B-EN. docx 4 Rev. 00 - 14/11/2012 -EN- CSK8B it is necessary, whenever the truck is loaded, try manually the free rotation of ALL spray arms (next photo) 2. 2 Hydraulic supply of trolley For the correct use of the CSK8B, another important step is to check the perfect contact between the nozzle supply trolley and that of the machine. If necessary, adjust the height by screwing / unscrewing ONLY the nozzle trolley. In the next picture you can see how to raise the nozzle trolley 19390404000_CSK8B-EN. docx 5 Rev. 00 - 14/11/2012 -EN- CSK8B - 2. 3 Drying surgical instruments Some models of Washer-disinfectors, are provided with the drying system. You can identify by the central hole, internally of the washer chamber. [. . . ] +39 0522 8211 – Fax +39 0522 821 592 E-Mail: instruments@smeg. it – www. smeg-instruments. com 19390404000_CSK8B-EN. docx 7 Rev. [. . . ]